Contact: WANG
Mobile: +86-138 0199 7980
Email: klaus@jnewtech.com
Address: BoYang Road No.16, Building23-A-207
UV Concentration System
- Hightlight:
- COMS Spectrometer
- Xenon Light Source
- Cost Efficient
- Uy Measurement
For Email Quatation
klaus@jnewtech.com
- Introduction
Automatic concentration detector (UV xenon lamp + UV spectrometer + probe)
It can detect the concentration of liquids such as magnesium ions, potassium ions, sodium ions, chloride ions, 200-1100 spectrum range, and uses 180-480 for UV enhanced spectrometer
Fully automatic concentration analysis, support self-learning, support self-built concentration gradient curve
It combines CMOS UV spectrometer and xenon light source to achieve a cost-effective solution.
CMOS UV spectrometer
Combined xenon light source
CMOS detector from JPN
Hardware from USA/Germany
1. Integrated UV spectrometer system, which provides a complete solution, combining CMOS spectrometer with xenon lamp, convenient for users to measure
2. Stable UV measurement, high-performance xenon flash lamp, especially in the UV region, makes its measurement stable
3. Low cost, the whole integrated system reduces complexity, is a complete UV measurement solution.
UVS3 series highlights
CMOS detector performs well in UV ultraviolet segment
High sensitivity in UV region wavelength range 180-500nm, 200-400nm
Signal-to-noise ratio 800:1 *snr=(signal-noise)/deviation
Integrated xenon light source
Reduce system complexity
Cost-effective UV spectroscopy system
Supports extended SDK and secondary development
SMA905 optical interface
Complete UVS3 system measurement accessories, such as flow cell
USB interface
We also provide OEM module
Application
UV spectrophotometer
UV spectrum measurement
UV spectrometer
UV absorption spectroscopy
UV absorbance, absorbance measurement
Transmittance, transmittance coefficient measurement
Water quality, total nitrogen measurement
Absorbance calculation, concentration calculation
Using Lambert-Beer's law, the concentration or content of the liquid can be calculated
Protein concentration measurement
Quantification Analysis and measurement
Application example 1: UV absorption, concentration analysis
Test steps
Put a reference sample in the cuvette. The reference sample can be water or nothing. If nothing is put, it is the reference air. The cuvette must be a quartz cuvette because quartz cuvettes have less interference with the UV spectrum than plastic cuvettes
Put the cuvette in the cuvette holder, the software controls the xenon lamp to turn off, and collect the electronic noise of the spectral equipment
The software controls the xenon lamp to turn on and collect the spectrum of the reference signal
Put the liquid to be measured, such as protein, or the liquid to be measured in the cuvette
The software controls the xenon lamp to turn on and collect the spectrum of the measured object. The absorbance of the liquid relative to water A = - log10 (transmittance value)
According to the Lambert-Beer law, we can deduce the concentration of the liquid
Mathematical expression of the Beer-Lambert law
A=-log10(T transmittance)=Kbc
A is absorbance, T is transmittance, It is the ratio of the intensity of the outgoing light (I) to the intensity of the incident light (I0).
K is the molar absorption coefficient. It is related to the properties of the absorbing substance and the wavelength λ of the incident light.
c is the concentration of the absorbing substance, in mol/L, and b is the thickness of the absorbing layer, in. [b is often replaced by L, with the same meaning]
Note: Absorbance has no unit. It is a ratio, which is the light intensity measured without liquid divided by the light intensity measured after adding liquid
From the Beer-Lambert law, it is not difficult to find that if the liquid concentration is very low, then the absorbance value will be very high. Ordinary detection instruments have a detection range, that is, a range ratio between the maximum light intensity and the minimum light intensity (we call it dynamic range). Ordinary spectrophotometers generally can only reach 100 times the ratio of the strongest light to the weakest light, that is, the absorbance value is 2 = -log(0.01) Our SPM series spectrometers can achieve an absorbance of 3, which is 1000 times higher. Compared with traditional UV spectrophotometers, SPM spectrometers have a large dynamic range, which means that our ability to detect weak light is much stronger than ordinary spectrophotometers. This is conducive to the detection of low-concentration liquids. We can use a small amount of liquid and dilute it for measurement. This can save expensive liquid costs.
- Previous: 没有了!
- Next: Slope Optical Path Length System
Contact
Hotline
+86-138 0199 7980
Office hour
Monday to Saturday
Company Phone
13801997980
QR Code
